• SKU: DURR-CRPH100001
  • Brand: DÜRR NDT GmbH & Co

DÜRR CR Phantom

Test specimen for Computed Radiography systems. Features two D15 duplex wire IQI‘s, BAM Snail, T-Target, Low-Contrast IQI in accordance with ISO 16371-1, ASTM E2445. Dimensions...
REQUEST A QUOTE
Available in North America Only
  • Competitive Shipping Rates
  • Satisfaction Guaranteed
  • Secure Checkout
  • Description
  • features
Description

Test specimen for Computed Radiography systems. Features two D15 duplex wire IQI‘s, BAM Snail, T-Target, Low-Contrast IQI in accordance with ISO 16371-1, ASTM E2445.

  • Dimensions (H x W x D): 200 x 275 x 15 mm (7.9 x 10.8 x 0.59")
  • Scope of supply: CR Phantom in lined wooden case,
    test reports according to ISO 16371-1ASTM E 2445,
    Declaration of Conformity according to EN 45014,
    test procedure (download at: www.duerr-ndt.com/register)
features

  • T-target (brass)
    Laser beam jitter, MTF check, blooming (flare)
  • Duplex wire-type IQI 15D (hi-res)
    Basic Spatial Resolution (unsharpness)
  • BAM snail
    Central beam alignment
  • Line pair IQI type 53-b
    Line pair resolution
  • Measuring points
    Shading correction
  • Cassette positioning locator
    Positioning of cassette (imaging plate)
  • Homogeneous strip (aluminum)
    Scanner slippage, scanner shading
  • Lucite plate
    Carrier plate
  • cm/inch ruler
    Linearity check
  • Contrast sensitivity gauge
    Contrast sensitivity check